6.0 – LCD ERROR CODES
(continued)
STAGE #4 – STERILIZATION
(continued)
LCD MESSAGE
CHECK CASSETTE #7 (CYCLE FAULT #7)
The steam generator/boiler or cassette temperature has
dropped below its set point due to a steam leak.
THINGS TO CHECK
• Clean and lubricate the cassette seal – see page 10,
Figure 2.7 - Liquid Soap (RPI Part #RPS287). Replace
the Cassette Seal if necessary, (RPI Part #SCS001 fits
model 2000), and (RPI Part #SCS029 fits model 5000).
• Inspect and clean steam ports on probe bracket.
• If the cassette is hard to remove after venting takes
place: Solenoid valve plunger may be compromised by
debris or may be worn. Check the plunger inside the
flute/bonnet for proper movement. Replace Solenoid-
Plunger Kit (RPI Part #SCK003), if necessary. See page
41, Figure 9.4 - Solenoid Valve.
• Water level in the waste bottle may be full, or the ex-
haust path may have a restriction or blockage.
• Check exhaust tubing for kinks or pinch points. Note: RPI
offers an Exhaust Tubing Kit (RPI Part #SCK017) or Ex-
haust Tubing sold by the foot (RPI Part #RPT380).
• Check the cassette tray venturi plate for blockage - see
page 42, Figure 9.5 - Cassette.
• Check steam generator/boiler – see page 38, Figure
9.1 - Steam Generator/Boiler.
LCD MESSAGE
CHECK CASSETTE #8 (CYCLE FAULT #8)
There is a 5˚C temperature difference between the steam
generator/boiler and the cassette.
THINGS TO CHECK
• W
ater level in the waste bottle may be full, or the ex-
haust path may have a restriction or blockage.
• Check exhaust tubing for kinks or pinch points. Note: RPI
offers an Exhaust Tubing Kit (RPI Part #SCK017) or Ex-
haust Tubing sold by the foot (RPI Part #RPT380).
• Check the cassette tray venturi plate for blockage – see
page 42, Figure 9.5 - Cassette.
• Check the solenoid valve coil’s electrical performance -
see page 41, Figure 9.4 - Solenoid Valve.
• Check for debris in the orifice and port holes of the
solenoid valve.
CHECK CASSETTE #8 (CYCLE FAULT #8)
(continued)
THINGS TO CHECK
(continued)
• Solenoid valve plunger may be compromised by debris
or may be worn. Check the plunger inside the flute/bon-
net for proper movement. Replace Solenoid-Plunger Kit
(RPI Part #SCK003) if necessary. See page 41, Figure
9.4 - Solenoid Valve.
• If problem repeats, clean or calibrate steam generator/
boiler - see page 39, Figure 9.2 or page 40, Figure
9.3.
LCD MESSAGE
CYCLE FAULT #9……
Not Currently In Use
LCD MESSAGE
CYCLE FAULT #10
Temperature and pressure failing to decrease.
THINGS TO CHECK
• Water level in the bottle may be full, or the exhaust path
may have a restriction or blockage.
• Check exhaust tubing for kinks or pinch points. Note: RPI
offers an Exhaust Tubing Kit (RPI Part #SCK017) or Ex-
haust Tubing sold by the foot (RPI Part #RPT380).
• Check the cassette tray venturi plate for blockage –
see page 42, Figure 9.5 - Cassette.
• Check the solenoid valve coil’s electrical performance -
see page 41, Figure 9.4 - Solenoid Valve.
• Check for debris in the orifice and port holes of the so-
lenoid valve.
• Solenoid valve plunger may be compromised by debris
or may be worn. Check the plunger inside the flute/bon-
net for proper movement. Replace Solenoid-Plunger Kit
(RPI Part #SCK003) if necessary. See page 41, Figure
9.4 - Solenoid Valve.
For older units with an aluminum steam generator/boiler,
run a new cycle and view the LCD display. After comple-
tion of the "Pressurization Stage", The "Sterilization
Stage" should begin within 10 seconds. If it takes longer
than 10 seconds, the steam generator/boiler needs clean-
ing - see page 39 - Figure 9.2 "Cleaning the Steam
Generator/Boiler".
6.0 – LCD ERROR CODES
(continued)
STAGE #5 – VENTING
The sterilization cycle is completed and the vent solenoid valve opens to: 1) allow the pressure in the cassette to depressurize
via the exhaust path into the condenser waste bottle; and, 2) allow the steam to cool down safely.
LCD MESSAGE
CYCLE FAULT #11
Cassette temperature failed to drop within 60 seconds of
the end of a cycl
e during venting.
THINGS TO CHECK
• Check exhaust tubing for kinks or pinch points. Note: RPI
offers an Exhaust Tubing Kit (RPI Part #SCK017) or Ex-
haust Tubing sold by the foot (RPI Part #RPT380).
• Check the cassette tray venturi plate for blockage – see
page 42, Figure 9.5 - Cassette.
• Check the solenoid valve coil’s electrical performance -
see page 41, Figure 9.4 - Solenoid Valve.
• Check for debris in the orifice and ports holes of the so-
lenoid valve.
• Solenoid valve plunger may be compromised by debris
or may be worn. Check the plunger inside the flute/bon-
net for proper movement. Replace Solenoid-Plunger Kit
(RPI Part #SCK003) if necessary. See page 41, Figure
9.4 - Solenoid Valve.
LCD MESSAGE
CYCLE FAULT #12
Failure to measure temperature. NOTE: This can happen at
any stag
e.
THINGS TO CHECK
• Check for open thermocouple(s). At room temperature,
it should read ~10 Ω. Replace if necessary – Thermo-
couple (RPI Part #SCT030).
LCD MESSAGE
CYCLE FAULT #13……
Not Currently In Use
LCD MESSAGE
CYCLE FAULT #14
Applies to STATIM 2xx software only.
Steam generator/boiler temperature "ran a
way" (above
171˚C) in the Sterilization Stage.
THINGS TO CHECK
• A weak pump due to dirty mesh filters in the pump fit-
tings or dirty Water Filter (RPI Part #SCF034). Check
pump’s operation – see pages 6 and 7, 2.3 - Pump
Tester Bottle (RPI Part #SCK024). If pump passes test,
boiler must be replaced.
• Steam generator/boiler calibration is needed – see page
40, Figure 9.3 - Steam Generator/Boiler Calibration
Procedure.
LCD MESSAGE
CYCLE FAULT #15
The cassette temperature climbed ~3˚C above the set
point limit in the Sterilization Stage.
THINGS TO CHECK
•
Water level in the waste bottle may be full, or the ex-
haust path may have a restriction or blockage.
• Check exhaust tubing for kinks or pinch points. Note: RPI
offers an Exhaust Tubing Kit (RPI Part #SCK017) or Ex-
haust Tubing sold by the foot (RPI Part #RPT380).
• Check the cassette tray venturi plate for blockage – see
page 42, Figure 9.5 - Cassette.
• Check for debris in the orifice and port holes of the
solenoid valve.
• Solenoid valve plunger may be compromised by debris
or may be worn. Check the plunger inside the flute/bon-
net for proper movement. Replace Solenoid-Plunger Kit
(RPI Part #SCK003) if necessary. See page 41, Figure
9.4 - Solenoid Valve.
• Steam generator/boiler calibration is needed – see page
40, Figure 9.3 - Steam Generator/Boiler Calibration
Procedure.