Replacement Parts Industries, Inc. is pleased to present this valuable work tool that can help
save you and your customers time and money. Included in this booklet you will find Instructions
on how to use each of the tools in this Kit, plus diagrams, exploded views, PM check lists,
error code listings with troubleshooting service tips and a complete list of RPI parts to fit the
SciCan STATIM Cassette Autoclaves.
For use with these models only:
STATIM 2000, model 101102 (with S/N ending in "A")
STATIM 2000, model 121101
STATIM 5000, model 01-201103
RPI Field Service Smart Kit
®
(
RPI Part #SCK028
)
A Guide to Maintaining
The SciCan STATIM
®
Cassette Autoclave
Zobrazit stránku 0
1 2 3 4 5 6 ... 28 29

Shrnutí obsahu

Strany 1 - RPI Part #SCK028

CALL (800) 221-9723 or (818) 882-8611 • FAX (818) 882-7028• E-MAIL [email protected] • WEBSITE www.rpiparts.comReplacement Parts Indust

Strany 2 - TABLE OF CONTENTS

16 17OILEVELSTARTSTOPREMOVE THIS LABELSERIES STELIZERFOR USE WITH 5000LEAVE THIS LABEL ONSERIES STERILIZERFOR USE WITH 2000DAILY OPERATIONIf the quick

Strany 3 - FILL LINE

18 19Cassette Seal SCS001Compressor Filter SCF002Biological Filter SCF032PART RPI PART # STERILIZER PM TIPRefer to the “Installation and Monthly Maint

Strany 4 - (RPI PART #SCS021)

20 21Cassette Seal SCS029Air Compressor Filter SCF031Biological Filter SCF032PART RPI PART # STERILIZER PM TIPRefer to the “Installation and Monthly M

Strany 5 - FIGURE 2.3A

22 23STEP #1 - The cassette must be opened and cooled before removing the cassette seal. See Figure A.STEP #2 - Start with any of the (4) corners and

Strany 6 - FIGURE 2.5

24 255.0 – SERVICE TIPS1. CASSETTE DIAGNOSTICSIf water is leaking at the drain from the cassette bay (un-derside of autoclave), then do the following:

Strany 7

6.0 – LCD ERROR CODES (Note: Error Codes that are not currently used by Sci-Can are listed as "Not Currently In Use" or not listedin this bo

Strany 8 - FIGURE 2.10A

28 296.0 – LCD ERROR CODES (continued)STAGE #3 – PRESSURIZINGThe pressure increases as the cassette temperature rises to the factory set temperatures

Strany 9 - FIGURE 2.12

30 316.0 – LCD ERROR CODES (continued)STAGE #4 – STERILIZATION(continued)LCD MESSAGE CHECK CASSETTE #7 (CYCLE FAULT #7)The steam generator/boiler or

Strany 10 - STATIM 2000 Shown

32 336.0 – LCD ERROR CODES (continued)STAGE #5 – VENTING(continued)LCD MESSAGE CYCLE FAULT #16Steam generator/boiler temperature "ran away"

Strany 11

34 357.0 – LCD MESSAGESAPPLIES TO ANY STAGELCD MESSAGEGARBLED OR NO READ-OUT ON THE LCDTHINGS TO CHECK• Make sure that the cable from the LCD board (R

Strany 12

1FIGURE 1 – STATIM 2000 & 5000 KEYPAD SYMBOLS CHARTRubber Model 2000 Unwrapped Wrapped & Plastics Air Dry Start StopRubber Heavy DutyModel 500

Strany 13 - Minimum Fill Line.)

36 378.0 – TROUBLESHOOTING TIPSCONDITION:WATER QUALITY SENSOR TROUBLESHOOTINGApplies to the STATIM models with water quality sensoronly.THINGS TO CHEC

Strany 14 - 5.0 – SERVICE TIPS

FIGURE 9.2 - CLEANING THE STEAM GENERATOR/BOILERFIGURE 9.1 - STEAM GENERATOR/BOILER38 399.0 – REFERENCE BLOCKS • To check the thermal fuse continuity

Strany 15 - 6.0 – LCD ERROR CODES

FIGURE 9.3 - STEAM GENERATOR/BOILER CALIBRATION PROCEDUREFIGURE 9.4 - SOLENOID VALVE40 419.0 – REFERENCE BLOCKS (continued)Applies to STATIM 2000 and

Strany 16

FIGURE 9.6 - PUMPFIGURE 9.5 - CASSETTE42 439.0 – REFERENCE BLOCKS (continued)CASSETTE DIAGNOSTICS:If water is leaking at the drain from the cassette

Strany 17

FIGURE 9.7 - AIR COMPRESSORThe OEM recommends replacing the compressor filters every 6 months or every 500 cycles (whichever comes first).Flow44 459.0

Strany 18

46 472143658710 912 1114 13151719P1ThermocoupleMicroswitchSolenoidValveAir CompressorSteam Generator/BoilerInvensysWaterQualitySensorThermocoupleHeate

Strany 19 - 7.0 – LCD MESSAGES

48 49FIGURE 9.9 - PLUMBING DIAGRAM9.0 – REFERENCE BLOCKS (continued)Female Quick ConnectFitting (Drain Line)ReservoirPumpIn-LineWater FilterSteam Gen

Strany 20 - 8.0 – TROUBLESHOOTING TIPS

50 51ADC058 No OEM Part # Available CORD CLIP - 10/PKG • •CAF011 No OEM Part # Available RUBBER FOOT - 4/PKG • •RPB365 53-100518A 90º LIQUID TIGHT STR

Strany 21 - 9.0 – REFERENCE BLOCKS

52 53NOTES11.0 - RPI TOOLS & ACCESSORIES LIST(FOR PARTS TO FIT STATIM 2000 & 5000)RPA369 No OEM Part # Available THREADLOCKER 545 • •RPB792 No

Strany 22 - FIGURE 1 FIGURE 2 FIGURE 3

Parts mentioned in this booklet are manufactured by Replacement Parts Industries, Inc. to fit SciCan STATIM® Cassette Autoclave Models 2000 and 5000.

Strany 23

23THISUPSIDEUPSIDETHISTHISUPSTOPSTARTLEVEL290FILL LINEMINIMUMGOLRKNNINCUOI2.1 – Extension Test CableRPI Part #SCT026(see page 4)2.2 – Start SwitchRPI

Strany 24

4 52.1 - EXTENSION TEST CABLE(RPI PART #SCT026)When testing, troubleshooting or repairing the STATIM au-toclave, install the RPI Extension Test Cable.

Strany 25

6 72.3 - PUMP TESTER BOTTLE(RPI PART #SCK024)The RPI bottle has a “swivel fitting” on the lid that allowsyou to control the position and balance of th

Strany 26 - FIGURE 9.9 - PLUMBING DIAGRAM

8 92.4 - SPANNER NUT(RPI PART #RPT364)When removing the solenoid flute/bonnet, it is suggestedby the OEM to use a pair of pliers on the welded area of

Strany 27 - 10.0 - RPI PARTS LIST

10 112.6 - CASSETTE SEAL REMOVALTOOL (RPI PART #RPT372)The Cassette Seal Removal Tool was designed to help remove the cassette seal for either mainten

Strany 28 - STATIM 5000

12 13FIGURE 2.10BTIP - If realigning existingprobe bracket, loosen (DO NOTREMOVE) 4 mounting screwsand compression fitting con-nections to allow repos

Strany 29 - SCK028INS REV - 03/13

TDS114 152.11 – WATER QUALITY TESTER(RPI PART #RPT820)The Water Quality Tester (RPI Part #RPT820) is a meter de-signed to measure water quality. This