4. Recover the biological and / or chemical indicator and process further according to the
literature that accompanied the indicator.
At the first indication of a potential sterilization failure:
1. Do not process any more instruments until favourable test results have been returned.
2. Ensure the correct indicator type was chosen.
3. Ensure the cassette was not overloaded. Consult the earlier portion of this section for
proper loading instructions.
4. If the results do not change, do not process any more instruments within the
Statim and contact your SciCan dealer for further assistance.
Because the turnaround time for the 3M Attest™ is up to 48 hours, it is recommended
that the tests be conducted so that the incubation period occurs during a period of
planned downtime such a the last cycle before a weekend.
4.6 Instrument Weight Guide
NOTE: The above weights are to be used as reference only. For exact weights of your
instruments, consult the manufacturer’s specifications.
16
4. Instructions For Use
Instrument
Scissors
Dental scalers
Forceps
Dental handpiece
Wrapped instrument rack
Unwrapped instrument rack
Suction cannula
Plastic mouth mirror
Impression tray
Plastic x-ray positioning ring
Typical Instrument Weight
30 g / 0.96 oz
20 g / 0.64 oz
15 g / 0.48 oz
40 to 60 g / 1.29 to 1.92 oz
260 g / 8.35 oz
225 g / 7.23 oz
10 g / 0.32 oz
8 g / 0.25 oz
15 to 45 g / 0.48 to 1.45 oz
20 g / 0.64 oz